Calibration Calculations

Calibrated oxygen reading:
O

2RC

= C

0

+ C

1

× O

2RU

Where:
C

1

= (O

2

100%Sat)/(O

2RUS

× O

2RUZ

)

C

0

= –C

1

× O

2RUZ

Where:
O

2

100%Sat is the theoretical 100% saturation point

O

2RUS

is the uncalibrated reading at 100% saturation

O

2RUZ

is the uncalibrated reading at 0% saturation

References:
Standard Methods for the Examination of Water and Wastewater

, 20th Ed,

2008. 4500-0 C. Azide Modifi cation. American Public Health Association. USA.