12. Technical Data

74.4 mm (2.93 in.)

25.15 mm (.99 in.)

109.2 mm (4.3 in.)

26.4 mm (1.04 in.)

Dimensions

General

Polarization Source:

Signet 2650 Amperometric Electronics

Compatible Flow Cells

:

Signet 3-3610-1 (159 001 683)
Signet 3-3610-2 (159 001 684)
Signet 3-4630.392 (159 001 690)

Mounting

:

Signet DryLoc connection

Materials

:

CPVC

Free Chlorine (prior to Nov 1, 2012):

Membrane material: PVDF

Free Chlorine (after to Nov 1, 2012) and Chlorine Dioxide:

Membrane material: PTFE

Free Chlorine and Chlorine Dioxide:

O-ring material: FPM
Working electrode: Gold
Counter reference electrode: Silver halide

Wetted Materials:

PVC, PVDF or PTFE, FPM, Nylon, Silicone

Performance

Electrode

:

Repeatability: ± 0.08 ppm (mg/l) or 3% of selected
range whichever is less
Free Chlorine Slope: 10 to 60 nA/ppm (mg/l) @ 25 ºC
Chlorine Dioxide Slope: 40 to 200 nA/ppm (mg/l) @ 17 ºC
Response time, T90: < 2 minutes

System:

(including electronics and instrument)

Accuracy: < ±3% of electrode signal after calibration
Resolution: ≤ 0.5% of electrode range

Sensor Conditioning:

New, fi rst start-up: 4 hours maximum before calibration
Subsequent start-ups: 2 hours maximum

Temperature Element:

PT1000

Operational Ranges and Limits

Free Chlorine:

Range: 3-2630-1: 0 to 2 ppm (mg/L)

3-2630-2: 0 to 5 ppm (mg/L)

3-2630-3: 0 to 20 ppm (mg/L)

pH Operating Range: 5.5 pH to 8.5 pH

Chlorine Dioxide:

Range: 3-2632-1: 0 to 2 ppm (mg/L)
pH Operating Range: 4.0 pH to 11 pH

Maximum Operating Temperature:

0 °C to 45 °C (32 °F to 113 °F)

Maximum Operating Pressure:

Membrane: 0.48 bar @ 25 °C (7 psi @ 77 °F)

Flow Velocity Across Membrane Surface:

Minimum: 15 cm/s (0.49 ft/s)
Maximum: 30 cm/s (0.98 ft/s)

Free Chlorine Cross Sensitivity:

Chlorine Dioxide, ozone, bromine

Chlorine Dioxide Cross Sensitivity:

Free Chlorine, ozone

Chemical Compatibility:

< 50% ethanol/water, < 50% glycerol/water

Environmental Requirements

Storage Temperature (dry):

-10 ºC to 60 ºC (-4 ºF to 140 ºF)

System Temperature:

-10 ºC to 60 ºC (-4 ºF to 140 ºF)

Relative Humidity:

0 to 95% indoor/outdoor non-condensing to rated ambient

Standards and Approvals

CE, WEEE
Manufactured under ISO 9001
RoHS Compliant
China RoHS (Go to www.gfsignet.com for details)

Declaration of Conformity according to FCC Part 15

This device complies with Part 15 of the FCC rules.

Operation is subject to the following two conditions:

(1) This device may not cause harmful interference, and,

(2) This device must accept any interference received,

including interference that may cause undesired operation.

This manual is related to the following products: