263X Series Chlorine Electrodes

7. Installation

Align tab on electrode with notch on
fi tting. Insert until fully seated.

Lubricate O-rings with a non-petroleum
based, viscous lubricant (grease)
compatible with the system.

Do Not Use Lubricant or Sealing Tape on Threads.
Do Not Overtighten. Do Not Use Tools.

6.2. Disposal

The electrode is not recyclable. Dispose of properly according to local, state and federal guidelines.

Reconditioning Procedure Continued:
(Steps 1-10, See Previous Page)
11. Fill beaker with a 12 mm (½ inch) of the appropriate solution.
12. Position or suspend the sensor 0.2 mm to 12 mm (¼ in. to ½ in.) above the appropriate

solution. DO NOT SUBMERGE THE SENSOR. See Figure 2.

13. Apply power to the system.
14. Monitor the nA of the sensor (press the down arrow once on the 8630 transmitter).

The nA reading should start to rise. Response time and nA reading will depend upon the
temperature of the bleach.

15. Once the sensor's nA reading reaches approximately 300-360 nA allow the sensor to

remain in the beaker, suspended over the appropriate solution, for an additional 20 minutes.

• If sensor does not recover quickly, cover the beaker to avoid air contamination.

• Contact the factory for assistance (www.gfsignet.com).

16. After 20 minutes, remove the sensor and install it into the fl ow cell and

restore fl ow to the system.

17. Calibrate the sensor after the system has become stable.

6.1. Storage of the Sensor continued

8. Mounting Position - PVC Tee

Flow

Best

Flow Direction

Good

45º

or

or

45º

Figure 2