GF Signet 2764-2767 Differential DryLoc pH_ORP Electrodes User Manual
Page 2
2
Signet 275X, 276X, 2777X Series pH and ORP Electrodes
Specifi cations:
276X Series Differential Electrodes
277X Series Threaded Electrodes
General
Compatibility
Signet 2750 and 2760
Shipping Weight
0.25 kg (0.55 lbs.)
Operating Range
276X Series 0 to 14 pH, ± 1500 mV ORP
277X Series 0 to 14 pH, ± 1500 mV ORP
Reference Electrolyte
276X Series Buffered KNO
3
solution
277X Series KCI/AgCl gel with KNO
3
salt bridge
Reference Element
276X Series
pH glass electrode
277X Series
Ag/AgCl
Temperature Sensor
276X Series pH
3 KΩ, PT1000 RTD, or 300 Ω
277X Series pH
3 KΩ or PT1000 RTD
276X Series ORP
10 KΩ ID for Signet instruments;
PT1000 RTD or 300 Ω options for use
with other instruments
277X Series ORP
10 KΩ ID for Signet instruments
Temp response time
95%
< 1 min.
Wetted Materials
Body
Ryton
®
Reference junctions
PTFE
Sensing surface
Glass (pH), Platinum (ORP)
O-rings
FPM
Solution ground
Carbon graphite (276X series only)
Standards & Approvals
CE
Manufactured under ISO 9001 and ISO 14001
Operating Temperature
276X Series
0 °C to 95 °C (32 °F to 203 °F)
277X Series
0 °C to 85 °C (32 °F to 185 °F)
Storage Temperature
> 0 °C (32 °F)
Max Operating Pressure 6.89 bar (100 psi)
0 40
50
60
70
°C
90 100
80
40
60
80
100
4
2
6
6.89
psi
bar
20
10 20 30
32 122
°F
212
176
50 86
277X Series
0 °C to 85 °C @ 6.89 bar
(32 °F to 185 °F @ 100 psi)
pH/ORP Electrodes
276X Series and 277X Series
Pressure vs. Temperature
276X Series ONL
Y
Specifi cations:
275X Series Wet-Tap Electrodes
General
Compatibility
Signet 2750 and 2760
Shipping Weight
0.25 kg (0.55 lbs.)
Operating Range
0 to 14 pH, ± 2000 mV ORP
Reference Electrolyte
Solidifi ed Acrylamide Gel
Wet-Tap electrolyte 3.5 M KCl
Reference Element
Ag/AgCl
pH Temp Sensor
PT-1000 or 3K Balco
ORP Identifi cation
10 KΩ ID for Signet instruments
Temp response time,
Wet-Tap
438 s
ORP response time
Application dependent
Effi ciency
> 97% @ 25 °C (77 °F)
Impedance (pH)
< 150 MΩ @ 25 °C
Sodium Ion Error
< 0.05 pH in 0.1 M Na+ ion at 12.8 pH
Wetted Materials
Wet-Tap body
Glass or PAS (Poly Aryl Sulphone)
Wet-Tap Junctions
Porous PTFE
Sensing surface
Glass (pH), Platinum (ORP)
O-rings
FPM
Solution ground
Carbon graphite (276X series only)
Standards & Approvals
CE
Manufactured under ISO 9001 and ISO 14001
Operating Temperature
0 °C to 85 °C (32 °F to 185 °F)
Storage Temperature
0 °C to 85 °C (32 °F to 185 °F)
Maximum Temperature and Pressure rating
6.89 bar @ 0 °C to 65 °C (100 psi @ 32 °F to 149 °F)
4.00 bar @ 65 °C to 85 °C (58 psi @ 150 °F to 185 °F)
China RoHS (Go to www.gfsignet.com for details)
China RoHS (Go to www.gfsignet.com for details)
Ryton
®
is a registered trademark of Chevron Phillips Chemical