beautypg.com

Datamax-O'Neil E-Class Mark III Operator’s Manual User Manual

Page 33

background image

Chapter 3 – Printer Operation

27

3.7.2

Manual Calibration

The Manual Calibration can be performed using the NETira CT Utility (see Section 3.5) or using the front panel

buttons via the printer’s menu, see Section 4.4.

Manual Calibration provides dynamic readings, which can be helpful when using media with small position-

critical notches or marks.

Calibrate the Media Sensor using the steps below:


Step

Action

Displayed Message

Comment

A

Turn ON the printer.

Wait briefly for initialization to

complete.

B


Using the

button scroll to MENU

and press the

button.

Make sure MEDIA SETTINGS is
highlighted and press the

button.


Then using the

button, scroll to

SENSOR CALIBRATION.

(Printer Menu)

C


Press the

button to access

SENSOR CALIBRATION.

Make sure MANUAL CALIBRATION
is highlighted and press the

button.


Using the

button, scroll to YES

and then press the

button to

proceed.

Manual Calibration

No

>Yes

Select NO to abort this

procedure.

D

With no media installed and the
cover closed, press the

button.

Remove Label Stock

Press Enter

Yyy

This sets the empty value, where

“yyy” represents the current

sensor reading.

E

Load the media. Be sure the Media

Sensor is positioned correctly.

With the media installed and the

cover closed, press and hold the

button until at least 2 full labels

have been fed from the printer and

then release.

Please Reload Media

Press/Hold Enter Key

yyy

Unless otherwise noted, do not

move the Media Sensor after this

step.

F

Observe the calibration result.

Calibration Complete

Calibration was successful.

If WARNING LOW BACKING is

displayed, calibration was still

successful.

G


Press the

repeatedly

to exit

the menu, select YES at the SAVE

CHANGES prompt and then press
the

button.

The printer is now ready for use.