2 positive transition filter, 3 negative transition filter, 4 event register – B&K Precision 4078 - Manual User Manual
Page 80: 5 event enable register
:STATus:QUEstionable:CONDition?
This query is used to read the condition register.
Command Type: Query only
Query
Syntax:
:STATus:QUES:COND?
Response:
NR1
4.13.5.4.2 Positive Transition Filter
:STAT:QUES:PTR
This command is used to set and query the value of the positive transition filter.
Arguments
Type:
NRf
Range:
0 to 131,072. Non integer arguments are rounded before execution.
Command Type: Setting or Query
Setting
Syntax:
:STAT:QUES:PTR
Examples:
:STAT:QUES:PTR 2048
Query
Syntax:
:STAT:QUES:PTR?
Response:
NR1
4.13.5.4.3 Negative Transition Filter
:STAT:QUES:NTR
This command is used to set and query the value of the negative transition filter.
Arguments
Type:
NRf
Range:
0 to 131,072. Non integer arguments are rounded before execution
Command Type: Setting or Query
Setting
Syntax:
:STAT:QUES:NTR
Examples:
:STAT:QUES:NTR 2048
Query
Syntax:
:STAT:QUES:NTR?
Response:
NR1
4.13.5.4.4 Event Register
:STAT:QUES:EVENt?
This query is used to read the event register. Reading the register clears it.
Command Type: Query only
Query
Syntax:
:STATus:QUES:EVEN?
Response:
NR1
4.13.5.4.5 Event Enable Register
:STAT:QUES:ENABle