Syntax:
[:SOURce:] PULse: WIDth?[
Response:
NR3
4.13.1.12.3 PULse EDGe
This command sets both rising and falling edge of the pulse to the specified value.
Arguments
Type:
Numeric
Units:
S, mS, uS, nS
Range:
100 nS minimum; maximum defined by period and width (see note above)
Rounding:
4 digits
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:] PULse:EDGe
[:SOURce:] PULse:EDGe
Examples:
[:SOURce:] PULse:EDGe 500NS
Query
Syntax:
[:SOURce:] PULse:EDGe?[
Response:
NR3
4.13.1.12.4 PULse RISe
This command sets rising edge of the pulse to the specified value.
Arguments
Type:
Numeric
Units:
S, mS, uS, nS
Range:
100 nS minimum; maximum defined by period and width (see note above)
Rounding:
4 digits
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:] PULse:RISe
[:SOURce:] PULse:RISe
Examples:
[:SOURce:] PULse:RISe 500NS
Query
Syntax:
[:SOURce:] PULse:RISe?[
Response:
NR3
4.13.1.12.5 PULse FALl
This command sets falling edge of the pulse to the specified value.
Arguments
Type:
Numeric
Units:
S, mS, uS, nS
Range:
100 nS minimum; maximum defined by period and width (see note above)
Rounding:
4 digits
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:] PULse:FALl
[:SOURce:] PULse:FALl
Examples:
[:SOURce:] PULse:FALl 500NS
Query
Syntax:
[:SOURce:] PULse:FALl?[
Response:
NR3
4.13.1.13 Duty Cycle
:SOURce:DCYCle