3 sweep time, 4 sweep start, 5 sweep stop – B&K Precision 4078 - Manual User Manual
Page 65

4.13.1.10.3 Sweep TIME
This command sets the time for one complete sweep:
Arguments
Type:
Numeric
Units:
S, mS, uS, nS
Range:
20mS to 500S
Rounding:
to 4 digits
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:]SWEEP:TIME
[:SOURce:]SWEEP:TIME
Examples:
SWEEP:TIME 50MS
Query
Syntax:
[:SOURce:]SWEEP:TIME?[
Response:
NR3
4.13.1.10.4 Sweep STARt
This command sets the start frequency of the sweep:
Arguments
Type:
Numeric.
Units:
MHz, KHz, Hz (default)
Range:
Dependent on the frequency range of the current function.
Rounding:
The value is rounded to 4 digits.
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:]SWEEP:STARt
[:SOURce:]SWEEP:STARt
Examples:
SWEEP:STARt 5KHZ
SWEEP:STARt 5E3
SWEEP:STARt MAXIMUM
SWEEP:STARt MIN
Query
Syntax:
[:SOURce:]SWEEP:STARt?[
Examples:
SWEEP:STARt ?
SWEEP:STARt ? MAX
Response: NR3
4.13.1.10.5 Sweep STOP
This command sets the stop frequency of the sweep:
Arguments
Type:
Numeric.
Units:
MHz, KHz, Hz (default)
Range:
Dependent on the frequency range of the current function.
Rounding:
The value is rounded to 4 digits.
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:]SWEEP:STOP
[:SOURce:]SWEEP:STOP
Examples:
SWEEP:STOP 5KHZ
SWEEP:STOP 5E3
SWEEP:STOP MAXIMUM
SWEEP:STOP MIN