B&K Precision 4078 - Manual User Manual
Page 101

95
Performance Tests Results
Setting
Minimum reading
Maximum reading
Reading
Frequency DDS
1 MHz
999980 Hz
1000020 Hz
Frequency ARB
1 MHz
999950 Hz
1000050 Hz
Sine Amplitude
10Vp-p
3.499V
3.572V
5Vp-p
1.749V
1.786V
3Vp-p
1.049V
1.072V
1Vp-p
0.349V
0.358V
100mVp-p
34 mV
37 mV
50mVp-p
17 mV
19 mV
Square Amplitude
10Vp-p
4.949V
5.051V
5Vp-p
2.474V
2.526V
3Vp-p
1.484V
1.516V
1Vp-p
0.494V
0.51
100mVp-p
49 mV
52 mV
50mVp-p
24 mV
26 mV
ARB Amplitude
10Vp-p
3.499V
3.572V
5Vp-p
1.749V
1.786V
3Vp-p
1.049V
1.072V
1Vp-p
0.349V
0.358V
100mVp-p
34 mV
37 mV
50mVp-p
17 mV
19 mV
Offset Accuracy
+4V
3.950V
4.050V
+2V
1.970V
2.030V
0V
-10mV
+10mV
-2V
-1.970V
-2.030V
-4V
-3.950V
-4.050V
Sine Distortion
1 KHz
- 65 dB
Square Transition
1 MHz
12 ns
Variable Duty Cycle
20%
19.00%
21.00%
50%
49.00%
51.00%