13 marker length, 14 marker state, 15 save – B&K Precision 4078 - Manual User Manual
Page 77
Setting
Syntax:
:ARBitrary:MARKer[:ADDRess]
Examples:
:ARB:MARK 45
Query
Syntax:
:ARBitrary:MARKer[:ADDRess]?
Example:
:ARB:MARK?
Response:
Marker address in NR1 format.
Considerations: The marker is only output if its address is within the range of addresses currently being run.
4.13.4.13 Marker Length
:ARBitrary:MARKer:LENGth
This command is used to set the marker length. The maximum allowed length of marker is 4000. The marker length
is specified by appending a numeric value, to the MARKer keyword. Maximum MARKER Length 4000 points
Arguments
Type:
Numeric
Command Type: Setting or Query
Setting
Syntax:
:ARBitrary:MARKer:LENGth
Example:
:ARB:MARK:LENG 5
Query
Syntax:
:ARBitrary:MARKer:LENGth?
Response:
NR1
4.13.4.14 Marker State
:ARBitrary:MARKer:STATe
This command is used to enable or disable the marker.
Arguments
Type:
Boolean
Command Type: Setting or Query
Setting
Syntax:
:ARBitrary:MARKer:STATe
Query
Syntax:
:ARBitrary:MARKer:STATe?
Response:
0|1
4.13.4.15 Save
:ARBitrary:SAVe
This command is used to save all unsaved arbitrary waveform data into non-volatile memory..
Arguments
Type:
none
Command Type: Setting only
Setting
Syntax:
:ARBitrary: SAVe