11 phase adjust, 12 pulse setting, 1 pulse period – B&K Precision 4078 - Manual User Manual
Page 66: 2 pulse width

Query
Syntax:
[:SOURce:]SWEEP:STOP?[
Examples:
SWEEP:STOP ?
SWEEP:STOP ? MAX
Response:
NR3
4.13.1.11 PHASe adjust
This command controls the phase adjustment
Arguments
Type:
Numeric
Units:
one (degrees implied)
Range:
-180 to +180 (other values converted to this range)
Rounding:
to integer
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:]PHASe
[:SOURce:]PHASe
Examples:
[:SOURce:]PHASe 500NS
Query
Syntax:
[:SOURce:]PHASe?[
Response: NR3
4.13.1.12 PULSe setting
The following commands control the pulse function:
Note that width +0.6*(rise + fall) < period in order to have valid values. Width < Period – 10 ns
4.13.1.12.1 PULSe PERiod
This command sets the pulse period to the specified value.
NOTE: This also sets the pulse frequency.
Arguments
Type:
Numeric
Units:
S, mS, uS, nS
Range: 100nS-1000S
Rounding:
4 digits
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:] PULse: PERiod
[:SOURce:] PULse: PERiod
Examples:
[:SOURce:] PULse: PERiod 500NS
Query
Syntax:
[:SOURce:] PULse: PERiod?[
Response:
NR3
4.13.1.12.2 PULse WIDth
This command pulse width to the specified value.
Arguments
Type:
Numeric
Units:
S, mS, uS, nS
Range:
20 nS minimum; maximum defined by period and transition (see note above)
Rounding:
4 digits
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:]PULse:WIDth
[:SOURce:] PULse:WIDth
Examples:
[:SOURce:] PULse:WIDth 500NS
Query