5 fsk source, 10 sweep control, 1 sweep state – B&K Precision 4078 - Manual User Manual
Page 64: 2 sweep spacing
Examples:
FSK:RATE 5KHZ
FSK:RATE 5E3
FSK:RATE MAXIMUM
FSK:RATE MIN
Query
Syntax:
[:SOURce]:FSK:RATE ?[
Examples:
FSK:RATE ?
FSK:RATE ? MAX
Response:
NR3
4.13.1.9.5 FSK SOURce
This command selects the FSK modulation source as either internal (then the above settings are effective) or external (and
then the external waveform determines the frequency of modulation).
Arguments
Type:
Character
Options:
INTernal, EXTernal
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:] FSK:SOURce
Examples:
FSK:SOUR INT
FSK:SOUR EXT
Query
Syntax:
[:SOURce]:FSK:SOURce?
Response:
INT|EXT
4.13.1.10 Sweep control
The following commands control the sweep functionality:
4.13.1.10.1 Sweep STATe
This command activates or deactivates sweep:
Arguments
Type:
Boolean
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:]SWEEP[:STATe]
Examples:
SWEEP:STAT ON
SWEEP OFF
Query
Syntax:
[:SOURce:]SWEEP[:STATe]?
Response:
0|1
4.13.1.10.2 Sweep SPACing
This command sets the sweep spacing as either LINear or LOGarithmic:
Arguments
Type:
Character
Options:
Linear, Logarithmic
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:] SWEEP:SPACing
Examples:
[:SOURce:] SWEEP:SPACing LIN
SWEEP:SPAC LOG
Query
Syntax:
[:SOURce:] SWEEP:SPACing ?
Response:
LIN|LOG