36585-series automatic calibrators (autocal) – Atec Anritsu-MS4640B User Manual
Page 41

VectorStar™ MS4640B Series
36585-Series Automatic Calibrators (AutoCal)
MS4640B Series TDS
PN: 11410-00611 Rev. A
41
36585-Series Automatic Calibrators (AutoCal)
The 36585-Series Precision Automatic Calibrator (AutoCal) Module provides industry-leading performance in corrected characteristics using
over-determined algorithms, and transferring characteristics from a highly accurate LRL type calibration. The resulting accuracies will even out perform a
Sliding Load SOLT calibration. In order to remove the effects of matched adapters, the Precision 36585-Series AutoCal comes in a variety of connector
gender types (m-m, f-f, and m-f). Adapter Removal Calibration routine is still available in the VectorStar software. With coverage from 70 kHz to 70 GHz,
the 36585-series Precision AutoCal offers not only the fastest and most reliable calibration, but also the most accurate broadband coaxial VNA calibration
method.
AutoCal General and Environmental
36581-Series Dimensions
65 mm H x 155 mm W x 90 mm D body (excluding connectors)
36585-Series Dimensions
42 mm H x 64 mm W x 140 mm D body (excluding connectors)
Control
Serial RS-232 control by the VNA via supplied 9-pin D-Sub cable
(allowing forward-compatibility to legacy AutoCAL)
Power
DC powered via supplied universal 110/220 V AC/DC adapter
(with enough power to maintain optimum stability)
Operating Temperature
18 to 28 °C
Storage Temperature
–20 to 70 °C
Relative Humidity
5 % to 95 % at 40 °C, Non-condensing
EMI
Conforms to and meets the requirements of:
EMC Directive
2004/108/EC
Low Voltage Directive
2006/95/EC
Emissions
EN55011:2009+A1:2010 Group 1 Class A
Immunity
EN 61000-4-2-2009, 4 kV CD, 8 kV AD
EN 61000-4-3:2006+A2:2010, 3 V/m
EN 61000-4-4:2004, 0.5 kV S-L, 1 kV P-L
EN 61000-4-5:2006, 0.5 kV S-L, 1 kV L-E
EN 61000-4-6:2009, 3 V
EN 61000-4-11:2004, 100 % @ 20 ms
36585V Series Precision AutoCal Module
36585 Series Precision AutoCal Calibration Kit
Description
Additional Information
Part Number
Precision AutoCal, K
70 kHz to 40 GHz, 2-port
K (male) to K (male)
36585K-2M
K (female) to K (female)
36585K-2F
K (male) to K (female)
36585K-2MF
Precision AutoCal, V
70 kHz to 70 GHz, 2-port
V (male) to V (male)
36585V-2M
V (female) to V (female)
36585V-2F
V (male) to V (female)
36585V-2MF