beautypg.com

For reflection uncertainties, it is assumed that s – Atec Anritsu-MS4640B User Manual

Page 21

background image

VectorStar™ MS4640B Series

MS4645B 50 GHz / MS4647B 70 GHz VNA System Performance

MS4640B Series TDS

PN: 11410-00611 Rev. A

21

MS4645B/MS4647B Measurement Uncertainties (Reflection)

The graphs give measurement uncertainties after the above calibration. The errors are a worst-case contribution of residual directivity, load and

source match, frequency response and isolation, network analyzer dynamic accuracy, and connector repeatability. 10 Hz IF Bandwidth is used. For

transmission uncertainties, it is assumed that S

11

= S

22

= 0. For reflection uncertainties, it is assumed that S

21

= S

12

= 0. All calibrations and

measurements were performed at 0 dBm or default port power, whichever is less. For other conditions, please use our free Exact Uncertainty

Calculator software, available for download from the Anritsu web site at www.anritsu.com.

Reflection Magnitude Uncertainty (1 of 2)

MS4647B with SOLT Calibration using 3654D Cal Kit

0.1

1

10

-40

-35

-30

-25

-20

-15

-10

-5

0

Device Reflection (dB)

Uncertainty (dB)

70 - 300 kHz

0.3 - 10 MHz

10 - 200 MHz

0.2 - 2.5 GHz

2.5 - 20 GHz

Reflection Magnitude Uncertainty (2 of 2)

MS4647B with SOLT Calibration using 3654D Cal Kit

0.1

1

10

100

-40

-35

-30

-25

-20

-15

-10

-5

0

Device Reflection (dB)

Uncertainty (dB)

20 - 40 GHz

40 - 65 GHz

65 - 67 GHz

67 - 70 GHz

Reflection Phase Uncertainty (1 of 2)

MS4647B with SOLT Calibration using 3654D Cal Kit

1

10

100

-40

-35

-30

-25

-20

-15

-10

-5

0

Device Reflection (dB)

Uncert

ainty (degrees)

70 - 300 kHz

0.3 - 10 MHz

10 - 200 MHz

0.2 - 2.5 GHz

2.5 - 20 GHz

Reflection Phase Uncertainty (2 of 2)

MS464B with SOLT Calibration using 3654D Cal Kit

1

10

100

-40

-35

-30

-25

-20

-15

-10

-5

0

Device Reflection (dB)

Uncertainty (degrees)

20 - 40 GHz

40 - 65 GHz

65 - 67 GHz

67 - 70 GHz