MS4640B Series TDS

Measurement Times

VectorStar™ MS4640B Series

Measurement Times

Measurement times include sweep time, and band-switching time, in single channel mode. Typical.
20 μs/point is achieved in true swept mode, with 25,000 points, with ALC turned on for level accuracy, with display turned-on for tuning purposes, with

locking turned-on for frequency accuracy and repeatability, with correction turned on to meet published residual specifications, and over the full span of

the product with all band-switch points to fully characterize a device.

Measurement Time (ms), SYNTHESIZED Sweep, Display ON and ALC ON

Measurement Time (ms) vs. Noise Floor (dBm), SYNTHESIZED Sweep, Display ON and ALC ON

Calibration

Sweep Width

IFBW

Measurement Time

401

Points

1,601

Points

25,000

Points

100,000

Points

Uncorrected or

1-port calibration

Narrow (≤ 1 GHz span

without band-switch points)

1 MHz

14

40

510

2,200

30 kHz

22

90

1,230

4,900

1 kHz

380

1,600

25,000

100,000

Wide (70 GHz span)

1 MHz

50

70

570

2,300

30 kHz

67

120

1,300

5,000

1 kHz

420

1,670

25,000

100,000

2-port calibration (per sweep)

Narrow (≤ 1 GHz span without

band-switch points)

1 MHz

14

40

510

2,200

30 kHz

22

90

1,230

4,900

1 kHz

400

1,610

25,000

100,000

Wide (70 GHz span)

1 MHz

50

70

570

2,300

30 kHz

67

120

1,300

5,000

1 kHz

420

1,670

25,000

100,000

Calibration

Full Band

Sweep

Measurement Time

1,601 Points

Achieved Noise Floor at

Maximum Frequency (dBm)

IFBW

(kHz)

2-port calibration (per sweep)

MS4642B

90

–85

100

190

–95

10

MS4644B

95

–80

100

190

–90

10

MS4645/47B

100

–75

100

190

–85

10