beautypg.com

Atec Anritsu-MS4640B User Manual

Page 24

background image

24

PN: 11410-00611 Rev. A

MS4640B Series TDS

MS4645B 50 GHz / MS4647B 70 GHz VNA System Performance

VectorStar™ MS4640B Series

MS4645B/MS4647B VNAs – 12-Term – 36585V V AutoCal

MS4645B 50 GHz and MS4647B 70 GHz VNAs, with 12-term Calibration, using the 36585V V AutoCal.

MS4645B/MS4647B Measurement Uncertainties (Transmission)

The graphs give measurement uncertainties after the above calibration. The errors are a worst-case contribution of residual directivity, load and

source match, frequency response and isolation, network analyzer dynamic accuracy, and connector repeatability. 10 Hz IF Bandwidth is used.
For transmission uncertainties, it is assumed that S

11

= S

22

= 0. For reflection uncertainties, it is assumed that S

21

= S

12

= 0. All calibrations and

measurements were performed at 0 dBm or default port power, whichever is less. For other conditions, please use our free Exact Uncertainty

Calculator software, available for download from the Anritsu web site at www.anritsu.com.

Frequency Range

(GHz)

Directivity

(dB)

Source Match

(dB)

Load Match

a

(dB)

a. Since Residual Load Match is limited by Residual Directivity and the user test port cable, it can only be specified as Residual Directivity. For

practical considerations, derate it by approximately 8 dB for a 3670 series test port cable, to compensate for effects such as match,

repeatability, bend radius, and similar parameters.

Reflection Tracking

(dB)

Transmission Tracking

(dB)

70 kHz to 0.01

b

b. Typical performance below 2 MHz.

> 40

> 40

> 40

± 0.10

± 0.10

0.01 to 2.5

> 43

> 47

> 43

± 0.05

± 0.03

2.5 to 20

> 50

> 47

> 50

± 0.09

± 0.03

20 to 40

> 48

> 47

> 48

± 0.14

± 0.07

40 to 65

> 43

> 45

> 43

± 0.17

c

c. ± 0.25 dB from 51 to 55 GHz.

± 0.10

65 to 67

> 43

> 45

> 43

± 0.17

± 0.10

67 to 70

> 42

> 40

> 42

± 0.30

± 0.12

Transmission Magnitude Uncertainty (1 of 2)

MS4647B with a 12-term Calibration using the 36585V AutoCal

0.01

0.1

1

10

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Device Transmission (dB)

Uncertainty (dB)

70 - 300 kHz

0.3 - 10 MHz

10 - 200 MHz

0.2 - 2.5 GHz

2.5 - 20 GHz

Transmission Magnitude Uncertainty (2 of 2)

MS4647B with a 12-term Calibration using the 36585V AutoCal

0.1

1

10

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Device Transmission (dB)

Uncertainty (dB)

20 - 40 GHz

40 - 65 GHz

65 - 67 GHz

67 - 70 GHz

Transmission Phase Uncertainty (1 of 2)

MS4647B with a 12-term Calibration using the 36585V AutoCal

0.1

1

10

100

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Device Transmission (dB)

Uncertainty (degrees)

70 - 300 kHz

0.3 - 10 MHz

10 - 200 MHz

0.2 - 2.5 GHz

2.5 - 20 GHz

Transmission Phase Uncertainty (2 of 2)

MS4647B with a 12-term Calibration using the 36585V AutoCal

0.1

1

10

100

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Device Transmission (dB)

Uncertainty (degrees)

20 - 40 GHz

40 - 65 GHz

65 - 67 GHz

67 - 70 GHz