Page 89

background image

Multi 3500i

What to do if...

87

Measured value too

high

Sensor symbol flashes

6.4

Conductivity measurement

Conductivity

measuring cell

was not recognized

Error message, Error

(invalid

calibration)

Sensor symbol flashes

Cause

Remedy

– High amount of dissolved

substances

– Correct solubility function

using the salinity equivalent

– Air bubbles bump on the

membrane with high velocity

– Avoid direct flow to the

membrane

– The carbon dioxide pressure is

too high (> 1 bar)

– Measuring not possible

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Measuring cell not connected

– Connect measuring cell

– Cable broken

– Replace measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and replace it if

necessary

– Unsuitable control standard

– Use control standard,

0.01 mol/l KCl

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system