5 ph calibration, Ph calibration – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 32

background image

Operation

Multi 3500i

30

4.4.5

pH calibration

Why calibrate?

pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is dis-
played. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the measuring instru-
ment. Thus, you should calibrate at regular intervals.

When to calibrate?

z After connecting another electrode
z When the sensor symbol flashes:

– after the calibration interval has expired

– after a voltage interruption (empty batteries)

Buffer sets for

calibration

You can use the buffer sets quoted in the table for an automatic calibra-
tion. The pH values are valid for the specified temperature values. The
temperature dependence of the pH values is taken into account during
calibration.

Note
The buffers are selected in the sensor menu (setting, Calibration type,
see section 4.4.4).

Buffer set

Name on the
display

pH values at
25 °C

Technical buffer solutions

TEC

2.00
4.01
7.00

10.01

NIST/DIN buffer solutions

NIST/DIN

1.679
4.006
6.865
9.180
12.454

(user-defined single-point or two-
point calibration)

ConCal

pH 7.0 ± 0.5
and any other
buffer solution