Lists

Multi 3500i

100

Salinity

The absolute salinity S

A

of seawater corresponds to the relationship

of the mass of dissolved salts to the mass of the solution (in g/Kg). In
practice, this dimension cannot be measured directly. Therefore, the
practical salinity is used for oceanographic monitoring. It is deter-
mined by measuring the electrical conductivity.

Salt content

General designation for the quantity of salt dissolved in water.

Setting the

temperature

compensation

Name of a function that considers the temperature influence on the
measurement and converts it accordingly. Depending on the mea-
sured parameter to be determined, the temperature compensation
functions in different ways. For conductimetric measurements, the
measured value is converted to a defined reference temperature. For
potentiometric measurements, the slope value is adjusted to the tem-
perature of the test sample but the measured value is not converted.

Slope

The slope of a linear calibration function.

Slope (relative)

Designation used by WTW in the DO measuring technique. It ex-
presses the relationship of the slope value to the value of a theoretical
reference sensor of the same type of construction.

Standard solution

The standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.

TDS

Total dissolved solids

Temperature

coefficient

Value of the slope of a linear temperature function.

Temperature function

Name of a mathematical function expressing the temperature behav-
ior of a test sample, a probe or part of a probe.

Test sample

Designation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.