2 ise measurement, Ise measurement – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 87

background image

Multi 3500i

What to do if...

85

6.2

ISE measurement

Error message OFL

Obviously incorrect

measured values

Error message, Error

(invalid

calibration)

Warning [TempErr]

Warning [ISEErr]

Cause

Remedy

– Measuring range exceeded

– Dilute test sample

Cause

Remedy

– Electrode not connected

– Connect electrode

– Cable broken

– Replace electrode

Cause

Remedy

Ion-sensitive electrode:

– Moisture in the plug

– Dry plug

– Electrode obsolete

– Replace electrode

– Electrode unsuitable for the

range to be measured

– Use a suitable electrode

– Socket damp

– Dry socket

Calibration procedure:

– Wrong sequence of standards

for three point calibration

– Select correct sequence

– Calibration standards do not

have the correct temperature
(max.

±

2 °C temperature

difference)

– Adjust the temperature of

the calibration standards

Cause

Remedy

– Temperature difference

between measurement and
calibration greater than 2 K.

– Adjust the temperature of

the test sample

Cause

Remedy

– Electrode voltage outside

calibrated range

– Recalibrate