Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual
Page 42

Operation
Multi 3500i
40
Continuing for two-point
calibration
(Calibration type
ConCal)
10
Thoroughly rinse the electrode with distilled water.
11
Immerse the electrode in buffer solution 2.
12
Set the nominal buffer value with and
<
▲> <▼>. Then confirm the value with .
13
If the Set temperature menu item appears, measure and enter
the temperature of the buffer manually (measurement without
temperature sensor).
14
Select Continue with <
▲> <▼> and press .
The buffer is measured.
The measured value is checked for stability (AutoRead).
15
Wait for the end of the AutoRead measurement or Terminate
AutoRead and accept the calibration value with
.
The new calibration values are displayed as a message (
i
).
You have the following options:
z Accept the new calibration values with .
Subsequently, the calibration record is displayed and output
to the interface at the same time.
z To switch to the measured value display without accepting
the new calibration values, press
i Buffer value = 8.85
i U = -102.0 mV
i Temperature =
24 °C
Terminate AutoRead
pH