Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 42

background image

Operation

Multi 3500i

40

Continuing for two-point

calibration

(Calibration type

ConCal)

10

Thoroughly rinse the electrode with distilled water.

11

Immerse the electrode in buffer solution 2.

12

Set the nominal buffer value with

and
<

> <>. Then confirm the value with

.

13

If the Set temperature menu item appears, measure and enter
the temperature of the buffer manually (measurement without
temperature sensor).

14

Select Continue with <

> <> and press

.

The buffer is measured.
The measured value is checked for stability (AutoRead).

15

Wait for the end of the AutoRead measurement or Terminate
AutoRead
and accept the calibration value with

.
The new calibration values are displayed as a message (

i

).

You have the following options:
z Accept the new calibration values with

.

Subsequently, the calibration record is displayed and output
to the interface at the same time.

z To switch to the measured value display without accepting

the new calibration values, press or .

i Buffer value = 8.85
i U = -102.0 mV
i Temperature =

24 °C

Terminate AutoRead

pH