3 settings for ise measurements, 4 calibrating for ise measurements, Settings for ise measurements – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 45: Calibrating for ise measurements

background image

Multi 3500i

Operation

43

adjust the temperature of the standard and measuring solutions ac-
cordingly. If the temperature difference is greater the [TempErr] warn-
ing appears in the measured value display.

4.5.3

Settings for ISE measurements

Overview

The following settings are possible for ISE measurements:

z Calibration record (display)

Settings

The settings are made in the measuring menu of the ISE measure-
ment. To open the settings, activate the relevant measuring window in
the measured value display and press the

key short-
ly. After completing the settings, switch to the measured value display
with .

4.5.4

Calibrating for ISE measurements

Why calibrate?

Ion-selective electrodes age and are temperature-dependent. This
changes the slope. As a result, an inexact measured value is displayed.
Calibration determines the current value of the slope of the electrode
and stores it in the instrument.

Thus, you should calibrate before each measurement and at regular in-
tervals.

When to calibrate?

z Before any measurement if possible
z After connecting another ISE electrode
z When the sensor symbol flashes, e.g. after a voltage interruption

(empty batteries)

Menu item

Possible
setting

Description

Calibration / Calibra-
tion record

-

Displays the calibration
record of the last calibra-
tion.

Man. temperature

-20 ... +130 °C

Entry of the manually de-
termined temperature. For
measurements without
temperature sensor only.