Multi 3500i

Operation

33

Calibration evaluation

After calibrating, the measuring instrument automatically evaluates the
calibration. The asymmetry and slope are evaluated separately. The
worse evaluation of both is taken into account. The evaluation appears
on the display and in the calibration record.

Preparatory activities

Perform the following preparatory activities when you want to calibrate:

Display

Calibration
record

Asymmetry
[mV]

Slope
[mV/pH]

+++

-15 ... +15

-60.5 ... -58

++

-20 ... +20

-58 ... -57

+

-25 ... +25

-61 ... -60.5
or
-57 ... -56

-

-30 ... +30

-62 ... -61
or
-56 ... -50

Clean the electrode according
to the electrode operating
manual

Error

Error

< -30 or
> 30

< -62 or
> -50

Perform error elimination ac-
cording to chapter 6 W

HAT

TO

DO

IF

...

1

Connect the pH electrode to the measuring instrument.
The pH/ISE measuring window is displayed.

2

Keep the buffer solutions ready. Adjust the temperature of the
buffer solutions, or measure the current temperature, if you
measure without a temperature sensor.