Sample record:

Calibration evaluation

After the calibration, the measuring instrument evaluates the current
status of the sensor against the relative slope. The evaluation appears
on the display and in the calibration record. The relative slope has no
effect on the measuring accuracy. Low values indicate that the electro-
lyte will soon be depleted and the sensor will have to be regenerated.

03.11.03 07:14

CALIBRATION ConOx

03.11.03 07:12:58

Multi 3500i ser. no. 12345678

Cal. interval 14 d

Relative slope 0.97

Sensor +++

Display

Calibration
record

Relative slope

+++

S = 0.8 ... 1.25

++

S = 0.7 ... 0.8

+

S = 0.6 ... 0.7

Error

Error

S < 0.6 or S > 1.25

Perform error elimination ac-
cording to chapter 6 W

HAT

TO

DO