4 settings for ph and orp measurements, Settings for ph and orp measurements – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual
Page 31

Multi 3500i
Operation
29
4.4.4
Settings for pH and ORP measurements
Overview
The following settings are possible for pH and ORP measurements:
z Resolution
z Calibration interval
z Buffers for calibration
z Unit for slope
z Calibration record (display)
Settings
The settings are made in the measuring menu of the pH/ORP measure-
ment. To open the settings, activate the relevant measuring window in
the measured value display and press the key short-
ly. After completing the settings, switch to the measured value display
with
Menu item
Possible
setting
Description
Calibration /
Calibration inter-
val
1 ... 999 d
Calibration interval for the pH elec-
trode (in days).
The measuring instrument reminds
you to calibrate regularly by the
flashing sensor symbol in the mea-
suring window.
Calibration /
Calibration type
TEC
NIST/DIN
ConCal
Buffer sets to be used for pH cali-
bration.
For details, see section 4.4.5.
Calibration / Unit
for slope
mV/pH
%
Unit of the slope.
The % display refers to the Nernst
slope of
-59.16 mV/pH (100 x determined
slope/Nernst slope).
Calibration /
Calibration
record
-
Displays the calibration record of
the last calibration.
Man. tempera-
ture
-20 ...
+130 °C
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Reset
-
Resets all sensor settings to the
delivery condition (see
section 4.10.1).
High resolution
On
Off
Resolution of the pH display:
On = 0.001
Off = 0.01