Multi 3500i

Operation

29

4.4.4

Settings for pH and ORP measurements

Overview

The following settings are possible for pH and ORP measurements:

z Resolution
z Calibration interval
z Buffers for calibration
z Unit for slope
z Calibration record (display)

Settings

The settings are made in the measuring menu of the pH/ORP measure-
ment. To open the settings, activate the relevant measuring window in
the measured value display and press the

key short-
ly. After completing the settings, switch to the measured value display
with .

Menu item

Possible
setting

Description

Calibration /
Calibration inter-
val

1 ... 999 d

Calibration interval for the pH elec-
trode (in days).
The measuring instrument reminds
you to calibrate regularly by the
flashing sensor symbol in the mea-
suring window.

Calibration /
Calibration type

TEC
NIST/DIN
ConCal

Buffer sets to be used for pH cali-
bration.
For details, see section 4.4.5.

Calibration / Unit
for slope

mV/pH
%

Unit of the slope.
The % display refers to the Nernst
slope of
-59.16 mV/pH (100 x determined
slope/Nernst slope).

Calibration /
Calibration
record

-

Displays the calibration record of
the last calibration.

Man. tempera-
ture

-20 ...
+130 °C

Entry of the manually determined
temperature. For measurements
without temperature sensor only.

Reset

-

Resets all sensor settings to the
delivery condition (see
section 4.10.1).

High resolution

On
Off

Resolution of the pH display:
On = 0.001
Off = 0.01