Operation

Multi 3500i

78

4.10

Reset

You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.

4.10.1 Resetting the sensor settings

Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.

pH

The following settings for pH measurements are reset to the default
settings with the Reset function:

Oxi

The following settings for DO measurements are reset to the default
settings with the Reset function:

Setting

Default settings

Calibration type

TEC

Cal. interval

7 d

Unit for slope

mV/pH

Measured variable

pH

High resolution

On

Asymmetry

0 mV

Slope

-59.16 mV

Temperature, manual

25 °C

Setting

Default settings

Cal. interval

14 d

Comparison meas.

Off

Measured variable

DO concentration

Relative slope (S

Rel

)

1.00

Salinity (value)

0.0

Salinity (function)

Off