Multi 3500i

Operation

37

Continuing for three-

point calibration

(Calibration type TEC)

14

Thoroughly rinse the electrode with distilled water.

15

Immerse the electrode in buffer solution 3.

16

If necessary, measure the temperature of buffer 3 manually,
then enter and confirm it with <

> <> and

in the Set temperature setting.

17

Select Continue with <

> <> and press

.

The buffer is measured.
The measured value is checked for stability (AutoRead).

18

Wait for the end of the AutoRead measurement or Terminate
AutoRead
and accept the calibration value with

.
The new calibration values are displayed as a message (

i

).

You have the following options:
z Accept the new calibration values with

.

Subsequently, the calibration record is displayed and output
to the interface at the same time.

z To switch to the measured value display without accepting

the new calibration values, press or .

i Buffer value = 4.010
i U = 184.0 mV
i Temperature =

24.8 °C

Terminate AutoRead

pH