Multi 3500i

Operation

47

Continuing for two-point

calibration

11

Wait for the end of the AutoRead measurement or accept the
calibration value with

.
The calibration display for the next standard solution appears.

i Standard = 0.010 mg/l
i U = 0.5 mV

Terminate AutoRead

ISE

i #1 0.010 mg/l 24 °C
i Immerse sensor in std. 2

Set temperature:

24 °C

Continue

Set standard:

0.020 mg/l

ISE

12

Thoroughly rinse the electrode with distilled water.

13

Immerse the electrode in standard solution 2.

14

Select the Set standard setting with <

> <> and press

.

15

Set the concentration of the standard solution with <

> <>

and press

.

16

Measure the temperature of the standard solution using a ther-
mometer.

17

Select the Set temperature setting with <

> <> and press

.

18

Set the temperature with <

> <> and press

.

19

Select Continue with <

> <> and press

.

The standard solution is measured.
The measured value is checked for stability (AutoRead).