Multi 3500i

Operation

51

Note
The ConOx double sensor can perform the salinity correction automat-
ically. The conductivity module of the sensor measures the salt content
of the test sample simultaneously with the DO. The measuring instru-
ment takes the measured value into account.

AutoRead

(drift control)

The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured vari-
able flashes until a stable measured value is available.

Criteria

With identical measurement conditions, the following applies:

4.6.3

Settings for DO sensors

Overview

The following settings are possible for DO sensors:

z Salinity correction
z Salinity (salinity equivalent)
z Calibration interval
z Comparison measurement

Measuring mode

Reproducibility

Response time

DO concentration

better than 0.05 mg/l

> 10 seconds

DO saturation in-
dex

better than 0.6 %

> 10 seconds

DO partial pres-
sure

Better than 1.2 mbar

> 10 seconds