Multi 3500i

Operation

45

Note
The calibration record is automatically transmitted to the interface after
calibrating.

Sample record:

Calibration evaluation

After calibrating, the measuring instrument automatically evaluates the
calibration.

03.11.03 07:14

CALIBRATION ISE

03.11.03 07:12:01

Multi 3500i ser. no. 12345678

Standard 1 0.010 mg/l

Standard 2 0.020 mg/l

Voltage 1 0.0 mV 24.0 °C

Voltage 2 9.0 mV 24.0 °C

Slope 29.9 mV

Sensor +++

Display

Calibration
record

Magnitude of the slope [mV]

+++

50.0 ... 70.0 or 25.0 ... 35.0

Error

Error

< 50 or > 70
or
< 25 or > 35

Perform error elimination ac-
cording to chapter 6 W

HAT

TO

DO

IF

...