handylab multi 12

What to do if...

71

ba75371e03

11/2009

6.4

General errors

Display LoBat

Instrument does

not react

to keystroke

Display

to

Probe symbol

flashes

Message

StoFull

Cause

Remedy

– Batteries almost empty

– Exchange the batteries (see

section 5.1 M

AINTENANCE

)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Hold the > key
depressed and switch the
instrument on

Cause

Remedy

– Time-out of the interface

– Check that the instrument is

connected

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– All memory locations are full

– Output data storage and clear

data storage