4.7

Transmitting data

You have the following possibilities of transmitting data:

z One of the following options:

– With the AutoStore function (page 45), measured values are

periodically saved internally (save interval Int 1) and output on the
interface.

– With the Data transmission interval function (Int 2), measured values

are periodically output on the interface (see below).

AutoStore (Int 1) covers the Data transmission interval

(Int 2).

z Press the key

This manually triggers a data transmission of the current measured value
to the serial interface at any time - independently of the selected intervals.

z With the Output data storage function (page 47), calibration data or saved

measured values are output on the interface.

4.7.1 Data transmission interval (Int 2)

The interval to the data transmission (Int 2) determines the chronological in-
terval between automatic data transmissions. After the selected interval ex-
pires, the current data record is transmitted to the interface.

Setting the

Data transmission

interval

The default setting for the interval is OFF.
To start the data transmission, set an interval (5 s, 10 s, 30 s, 1 min, 5 min,
10 min, 15 min, 30 min, 60 min):

1

Press the key and hold it down.

2

Press the key. Int 2 appears on the display.

3

Set the required interval between the saving procedures with
<

> <>.

Store

Auto