9 reset, Reset, Ee section 4.9 r – Xylem handylab multi 12 User Manual

Page 61: Eset

background image

handylab multi 12

Operation

61

ba75371e03

11/2009

4.9

Reset

You can reset (initialize) the measurement parameters and the configuration
parameters separately from one another.

Measurement

parameters

The following measuring parameters (pH/O

2

/Cond InI) can be reset to the de-

fault condition:

pH

O

2

Conductivity

Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.

Configuration

parameters

The following configuration parameters (InI) are reset to the delivery status:

Measuring mode

pH

Asymmetry

0 mV

Slope -59.16

mV

Measuring mode

D. O. concentration

Relative slope

1.00

Salinity (value)

0.0

Salinity (function)

Off

Measuring mode

ӕ

Cell constant

0.475 cm

-1

(fixed)

Baud rate

4800

Display during the
pH calibration

Buffer nominal value

Interval 1
(automatically saved)

OFF

Interval 2
(for data transmission)

OFF