beautypg.com

Xylem handylab multi 12 User Manual

Page 32

background image

Operation

handylab multi 12

32

ba75371e03

11/2009

Probe evaluation

After the calibration, the measuring instrument evaluates the current status
of the probe against the relative slope. The evaluation appears on the dis-
play. The relative slope has no effect on the measuring accuracy. Low values
indicate that the electrolyte will soon be depleted and the probe will have to
be regenerated.

Starting the

calibration

Proceed as follows to calibrate the instrument:

Note
The sponge in the air calibration vessel must be moist (not wet).

Display

Relative slope

S = 0.8 ... 1.25

S = 0.7 ... 0.8

S = 0.6 ... 0.7

E3

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

S < 0.6 or S > 1.25

1

Perform the preparatory activities according to section 4.3.1.

2

Keep the air calibration vessel ready.

3

Put the D. O. probe into the air calibration vessel.

4

Press the key repeatedly until the calibration mode ap-
pears. The sensor symbol displays the evaluation of the last
calibration (or no sensor symbol in the delivery state or after the
measurement parameter has been reset).