Xylem handylab multi 12 User Manual
Page 10
Overview
handylab multi 12
10
ba75371e03
11/2009
Precision of
Oxi (± 1 digit)
Correction
functions of Oxi
Measuring ranges
of Conductivity
Precision of
Conductivity (± 1
digit)
Measuring range II
Resolution
0 ... 90.0l
0.1
0 ... 600
1
mg/l
%
Concentration
± 0.5 % of measured value
at ambient temperature of 5 °C ... 30 °C
Saturation
± 0.5 % of measured value
when measuring in the range of
± 10 K around the calibration temperature
Temperature
compensation
< 2 % at 0 ... 40 °C
T [°C]
± 0.1
Salinity correction
0 ... 70.0 SAL
Air pressure correction
Automatic through built-in pressure sensor
in the range 500 ... 1100 mbar
Measuring range
Resolution
ӕ
[µS/cm]
0 ... 1999
1
ӕ
[mS/cm]
0.00 ... 19.99
0.0 ... 199.9
0 ... 500
0.01
0.1
1
SAL
0.0 ... 70.0 accord-
ing to the IOT table
0.1
T [°C]
− 5.0 ... + 105.0
0.1
ӕ
Nonlinear compensation
:
Precision Temperature
of
material
under test
± 0.5 %
0 °C ... 35 °C
according to EN 27 888;
± 0.5 %
35 °C ... 50 °C
extended nLF function
SAL
Range 0.0 ... 42.0
Accuracy Sample
temperature
± 0.1
5 °C ... 25 °C
± 0.2
25 °C ... 30 °C
T [°C]
± 0.1