Overview

handylab multi 12

10

ba75371e03

11/2009

Precision of

Oxi (± 1 digit)

Correction

functions of Oxi

Measuring ranges

of Conductivity

Precision of

Conductivity (± 1

digit)

Measuring range II
Resolution

0 ... 90.0l
0.1

0 ... 600
1

mg/l

%

Concentration

± 0.5 % of measured value
at ambient temperature of 5 °C ... 30 °C

Saturation

± 0.5 % of measured value
when measuring in the range of
± 10 K around the calibration temperature

Temperature
compensation

< 2 % at 0 ... 40 °C

T [°C]

± 0.1

Salinity correction

0 ... 70.0 SAL

Air pressure correction

Automatic through built-in pressure sensor
in the range 500 ... 1100 mbar

Measuring range

Resolution

ӕ

[µS/cm]

0 ... 1999

1

ӕ

[mS/cm]

0.00 ... 19.99
0.0 ... 199.9
0 ... 500

0.01
0.1
1

SAL

0.0 ... 70.0 accord-
ing to the IOT table

0.1

T [°C]

− 5.0 ... + 105.0

0.1

ӕ

Nonlinear compensation

:

Precision Temperature

of

material

under test

± 0.5 %

0 °C ... 35 °C
according to EN 27 888;

± 0.5 %

35 °C ... 50 °C
extended nLF function

SAL

Range 0.0 ... 42.0
Accuracy Sample

temperature

± 0.1

5 °C ... 25 °C

± 0.2

25 °C ... 30 °C

T [°C]

± 0.1