2 pc/external printer (rs 232 interface), 3 remote control, Pc/external printer (rs232 interface) – Xylem handylab multi 12 User Manual
Page 56: Remote control, Section 4.7.2 bef
Operation
handylab multi 12
56
ba75371e03
11/2009
4.7.2
PC/external printer (RS232 interface)
Via the RS 232 interface, you can transmit the data to a PC or an external
printer.
Use the Z 395 (PC) or Z 391 (ext. printer) cable.
Warning
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.
Set up the following transmission data on the PC/printer:
Socket
assignment
4.7.3
Remote control
The handheld meter can be remotely controlled from a PC. This requires a
communication kit (communication cable + communication program) which is
available upon request. The instrument is then controlled via commands that
simulate keystrokes and request the current display contents.
Note
A more detailed description is provided within the scope of delivery of the
communication kit.
Baud rate
can be selected from: 1200, 2400, 4800, 9600
Handshake
RTS/CTS + Xon/Xoff
PC only:
Parity
none
Data bits
8
Stop bits
2
1 CTS
2 RxD
3 Ground
4 TxD
1
2
3
4
RS 232
REC