3 measuring the orp voltage, Measuring the orp voltage – Xylem handylab multi 12 User Manual
Page 21
handylab multi 12
Operation
21
ba75371e03
11/2009
Note
The current AutoRead measurement can be terminated at any time (accept-
ing the current value) by pressing
4.2.3
Measuring the ORP voltage
In conjunction with an ORP electrode, e.g. BlueLine 31Rx, the measuring in-
strument can measure the ORP voltage (U) of a solution.
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes
using a test solution.
3
Start AutoRead with
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4
If necessary, start the next AutoRead measurement with
5
To terminate the AutoRead function: Press the
1
Perform the preparatory activities according to section 4.2.1.
2
Submerse the Redox electrode in the sample.
3
Press the <
▲> <▼> keys until the U status display appears.
The ORP voltage (mV) of the sample appears on the display.
4
Wait for a stable measured value.
/pH