Measured value

too low

Measured value

too high

6.3

Conductivity system messages

Error message

OFL

Error message

E3

Cause

Remedy

– Insufficient flow

– Provide flow to the probe

Cause

Remedy

– High amount of dissolved

substances

– Correct solubility function

using the salinity equivalent

– Air bubbles bump on the

membrane with high velocity

– Avoid direct flow to the

membrane

– The carbon dioxide pressure is

too high (> 1 bar)

– Measuring not possible

Cause

Remedy

The measured value lies outside
the measuring range

– Measuring cell not connected

– Connect measuring cell

– Cable broken

– Replace measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and replace it if

necessary

– Unsuitable calibration

solution

– Check calibration solutions