beautypg.com

Al l c c – Xylem handylab multi 12 User Manual

Page 40

background image

Operation

handylab multi 12

40

ba75371e03

11/2009

Calibration

evaluation

After the calibration, the measuring instrument automatically evaluates the
current status of the calibration. The evaluation appears on the display.

Determining the

cell constant

You can determine the cell constant as follows:

Display

Cell constant [cm

-1

]

0.450 ... 0.500 cm

-1

E3

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

Outside the range
0.450 ... 0.500 cm

-1

1

Press the key repeatedly until the status display

ӕ

CAL CELL appears.

2

Press the key. The CAL display appears, as
well as
– the current, calibrated cell constant (with sensor symbol on

the display) or

– the fixed cell constant 0.475 1/cm (without sensor symbol

on the display). In this case, the measurement parameters
are initialized (see section 4.9 R

ESET

).

3

Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.

4

Press the

key.

The AutoRead measurement to determine the cell constant
starts. The AR display indicator flashes until a stable signal is
reached. The cell constant determined is displayed. The mea-
suring instrument automatically stores the cell constants.

AR

AR

RCL

RCL

O

O

UpH

UpH

Sal

Sal

Store

Store

Auto

Auto

AutoCal DIN

AutoCal DIN

AutoCal TEC

AutoCal TEC

LoBat

LoBat

ARng

ARng

Tref20

Tref20

Lin

Lin

nLF

nLF

Sal

Sal

%

%

mg/l

mg/l

mV/pH

mV/pH

cm

cm

°

°

°

°

TP

TP

m

m

S/

S/

cm

cm

F

F

C

C

%

%

/

/

K

K

M

M

TDS

TDS

1

1

S

S

mbar

mbar

Time

Baud

Day.Month No.

Ident

Year

Time

Baud

Day.Month No.

Ident

Year

5

2

0

0

7

7

4

4

A

A

L

L

C

C

cm

cm

1/

1/

Tref25

Tref25

Cal

Cal