4.4

Conductivity

4.4.1

General information

Caution
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.
The RS232 interface is not galvanically isolated.

Preparatory

activities

Perform the following preparatory activities when you want to measure:

Temperature probe

Only carry out measurements with a temperature probe. The temperature
probe is shown on the display by TP.

Temperature

compensation

The nonlinear temperature compensation is set fixed and is shown on the dis-
play by nLF.

Reference

temperature, Tref

The reference temperature (Tref) can be switched between 20 °C and 25 °C.
It appears on the display as Tref20 or Tref25. To switch over the reference
temperature, see S

WITCHING

OVER

THE

REFERENCE

TEMPERATURE

, page 59.

1

Connect a conductivity measuring cell to the measuring instru-
ment. The instrument automatically switches over to conductiv-
ity measurement (status display

ӕ

).

If the conductivity measuring cell is already connected, press
the key repeatedly until the status display

ӕ

or Sal ap-

pears.

2

Check the selected cell constant or calibrate the measuring in-
strument with the measuring cell (see below).

3

Check the selected cell constant (see below) or calibrate the
measuring instrument with the measuring cell (see section
4.4.5 on page 39)
.

4

Using <

> <>, toggle between the measuring modes, con-

ductivity (

ӕ

in

μS/cm) or salinity (SAL).