Oxi system messages

Error message

0FL

Error message

E3

Error message

E7

AR flashes

continuously

Cause

Remedy

Display range exceeded

Oxygen sensor

– Not connected

– Connect the probe

– Cable broken

– Replace probe

– Depleted

– Replace probe

– Short-circuit between gold and

lead electrode

– Clean probe and replace it if

necessary

Cause

Remedy

Invalid calibration

Oxygen sensor

– Electrolyte solution depleted

– Regenerate probe

– Membrane contaminated

– Clean membrane

– Electrode system poisoned

– Regenerate probe

– Worn out

– Replace probe

– broken

– Replace probe

Cause

Remedy

Membrane damaged

– Membrane damaged

– Regenerate probe

– Membrane head not screwed

on tight enough

– Screw membrane head tight

Cause

Remedy

No stable measured value

– Membrane contaminated

– Clean membrane