Operation

handylab multi 12

34

ba75371e03

11/2009

4.3.6

Salt content correction

A salt content correction is required in the oxygen concentration measure-
ment of samples with a salt content of more than 1 g/l. To do this, you have
to enter the salinity equivalent of the test sample (range 0.0 - 70.0) and to
switch on the salinity correction.

Note
With the handylab multi 12 handheld meter, you can measure the salinity.
How to proceed is described in section 4.4.3 M

EASURING

THE

SALINITY

on

page 37.

Entering the salt

content

Note
How to switch on the salt content correction is described on page 28.

1

Determine the salinity of the test sample (any method, see also
section 4.4.3 M

EASURING

THE

SALINITY

on page 37).

2

Press the key repeatedly until Sal appears on the dis-
play.

3

Enter the salt content with <

> <>.

4

Switch to the measuring mode with .

°