5 calibration intervals (int 3, int 4, int 5), Calibration intervals (int 3, int 4, int 5), T 4 n – Xylem handylab multi 12 User Manual
Page 42

Operation
handylab multi 12
42
ba75371e03
11/2009
4.5
Calibration intervals (Int 3, Int 4, Int 5)
For each measured variable, a time interval is stored. When it has expired,
you will be reminded to calibrate. After a calibration interval has expired, the
probe symbol of the relevant measured variable flashes. It is still possible to
measure. By calibrating the relevant probe, the function is reset and the in-
terval starts anew.
The following calibration intervals are set in the factory:
Setting the
calibration interval
You can change each of these intervals (1 ... 999 days):
Measured parameter
Designation
Default setting
pH/ORP
Int 3
7 days
Dissolved oxygen
Int 4
14 days
Conductivity
Int 5
180 days
1
Switch off the measuring instrument.
2
Press the
3
Press the
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configu-
ration.
4
Press the
gether with the required measured variable (pH, O
2
or
ӕ
) ap-
pears on the display. Example:
5
Set the required interval until the next calibration with <
▲>
<
▼>.
6
Confirm with
7
Switch to the measuring mode with
°
C
1
TP
ARng
mg/l
LoBat
Baud
Day.Month No.
Ident
Year
Sal
%
mbar
S
RCL
AR
Store
Auto
°
F
1/cm
/
K
%
TDS
UpH
Sal
mV/pH
S/
m
cm
cm
M
Lin
Tref25 Tref20
nLF
AutoCal DIN
AutoCal TEC
4
1
t 4
n
O
1
d
Time
Cal