handylab multi 12

Operation

19

ba75371e03

11/2009

Temperature

measurement in

pH measurements

You can perform pH measurements with or without a temperature probe as
well as with the temperature probe of an oxygen sensor or a conductivity
measuring cell. The measuring instrument recognizes which sensors are
connected and switches automatically to the correct mode for the tempera-
ture measurement.

The following cases are distinguishable.

* If you do not wish that, you can:

– either disconnect the 2nd sensor and use the manual temperature input

or

– use an electrode with a temperature probe.

If a temperature probe is connected, it is indicated on the display by TP.

Note
When calibrating without a temperature probe (no TP display indicator dis-
played), enter the current temperature of the respective buffer solution man-
ually using the <

> <> keys while keeping the key

depressed.

Temperature probe

Display

Mode

pH

Cond or
Oxi

yes

-

TP

Automatic with
pH temperature probe

yes

yes

TP

-

-

Manual

-

yes

TP
flashes

The temperature value of the
second probe (Cond or Oxi) in
the same sample is taken
over for the pH measurement*