Xylem handylab multi 12 User Manual
Page 19

handylab multi 12
Operation
19
ba75371e03
11/2009
Temperature
measurement in
pH measurements
You can perform pH measurements with or without a temperature probe as
well as with the temperature probe of an oxygen sensor or a conductivity
measuring cell. The measuring instrument recognizes which sensors are
connected and switches automatically to the correct mode for the tempera-
ture measurement.
The following cases are distinguishable.
* If you do not wish that, you can:
– either disconnect the 2nd sensor and use the manual temperature input
or
– use an electrode with a temperature probe.
If a temperature probe is connected, it is indicated on the display by TP.
Note
When calibrating without a temperature probe (no TP display indicator dis-
played), enter the current temperature of the respective buffer solution man-
ually using the <
▲> <▼> keys while keeping the
depressed.
Temperature probe
Display
Mode
pH
Cond or
Oxi
yes
-
TP
Automatic with
pH temperature probe
yes
yes
TP
-
-
Manual
-
yes
TP
flashes
The temperature value of the
second probe (Cond or Oxi) in
the same sample is taken
over for the pH measurement*