11/2009

4.3.2

Measuring the D. O. concentration

When measuring the concentration of test samples with a salt content of
more than 1 g/l, a salinity correction is required.

Note
How to enter the current salt content is described in section 4.3.6 S

ALT

CON

-

TENT

CORRECTION

on page 34.

Switching the salt content correction on or off, see below.

To measure the D. O. concentration with and without salt content correction,
proceed as follows:

Switching on/off

the salt content

correction

Proceed as follows to switch on the salt content correction:

1

Perform the preparatory activities according to section 4.3.1.

2

Immerse the D. O. probe in the test sample.

3

Press the <

> <> key repeatedly until the D. O. concentra-

tion in mg/l appears on the display.

RCL

cm

cm

M

m

S/

%

/pH

mV

pH

U

TDS

Tref20

Cal

8

S

1

Sal

1/cmSal

AutoCal DIN

Lin

Tref25

nLF

AutoCal TEC

Auto

Store

LoBat

AR

Time

Baud

Day.Month No.

Ident

Year

mbar

°

F

/

K

%

6

7

1

9 2

°

C

TP

7

O

mg/l

ARng

1

Perform the preparatory activities according to section 4.3.1.

2

Immerse the D. O. probe in the test sample.

3

While pressing the key, switch on the salt con-
tent correction with <

> . The SAL display indicator appears

on the display. The specified salt content is taken into consid-
eration during the measurement.