Xylem handylab multi 12 User Manual

Page 24

background image

Operation

handylab multi 12

24

ba75371e03

11/2009

Preparatory

activities

AutoCal TEC

For this procedure, use any two technical buffer solutions (pH values at
25 °C: 2.00 / 4.01 / 7.00 / 10.01).

Note
The calibration for pH 10.01 is optimized for the SI Analytics technical buffer
solution pH 10.01. Other buffer solutions can lead to an erroneous calibra-
tion. The correct buffer solutions are given in the SI Analytics catalog or in the
Internet.

Note
The buffer solutions are automatically recognized by the measuring instru-
ment. Depending on the instrument setting (see section 4.8 C

ONFIGURATION

),

the instrument displays the relevant buffer nominal value or the current elec-
trode voltage in mV.

Note
Skip the steps 2 and 7 if you use a pH electrode with temperature probe or
the temperature probe of a conductivity measuring cell or a D. O. probe.

Starting the

calibration

1

Connect the pH electrode to the measuring instrument.
If necessary, press the key repeatedly until the status dis-
play pH (pH measurement) or U (measurement of the ORP
voltage) appears.

2

Keep the buffer solutions ready.

3

Adjust the temperature of the solution and measure the current
temperature if the measurement is made without the use of a
temperature probe (the TP display indicator is missing from the
display).

1

Press the key. The Ct1 display and the function display
AutoCal TEC appears. The sensor symbol displays the evalu-
ation of the last calibration (or no sensor symbol in the delivery
state or after the measurement parameter has been reset).

/pH

mV

TDS

O

U

Tref20

Cal

8

S

1

Sal

%

mg/l

1/cmSal

AutoCal DIN

Lin

Tref25

nLF

RCL

Auto

Store

LoBat

ARng

AR

Time

Baud

Day.Month No.

Ident

Year

mbar

cm

M

°

F

/

K

%

S/

m

cm

8

4

2

t 1

°

C

TP

pH

C

AutoCal TEC