Xylem handylab multi 12 User Manual
Page 24
Operation
handylab multi 12
24
ba75371e03
11/2009
Preparatory
activities
AutoCal TEC
For this procedure, use any two technical buffer solutions (pH values at
25 °C: 2.00 / 4.01 / 7.00 / 10.01).
Note
The calibration for pH 10.01 is optimized for the SI Analytics technical buffer
solution pH 10.01. Other buffer solutions can lead to an erroneous calibra-
tion. The correct buffer solutions are given in the SI Analytics catalog or in the
Internet.
Note
The buffer solutions are automatically recognized by the measuring instru-
ment. Depending on the instrument setting (see section 4.8 C
the instrument displays the relevant buffer nominal value or the current elec-
trode voltage in mV.
Note
Skip the steps 2 and 7 if you use a pH electrode with temperature probe or
the temperature probe of a conductivity measuring cell or a D. O. probe.
Starting the
calibration
1
Connect the pH electrode to the measuring instrument.
If necessary, press the
play pH (pH measurement) or U (measurement of the ORP
voltage) appears.
2
Keep the buffer solutions ready.
3
Adjust the temperature of the solution and measure the current
temperature if the measurement is made without the use of a
temperature probe (the TP display indicator is missing from the
display).
1
Press the
AutoCal TEC appears. The sensor symbol displays the evalu-
ation of the last calibration (or no sensor symbol in the delivery
state or after the measurement parameter has been reset).
/pH
mV
TDS
O
U
Tref20
Cal
8
S
1
Sal
%
mg/l
1/cmSal
AutoCal DIN
Lin
Tref25
nLF
RCL
Auto
Store
LoBat
ARng
AR
Time
Baud
Day.Month No.
Ident
Year
mbar
cm
M
°
F
/
K
%
S/
m
cm
8
4
2
t 1
°
C
TP
pH
C
AutoCal TEC