4 autoread ar (drift control), Autoread ar (drift control) – Xylem handylab multi 12 User Manual
Page 30
Operation
handylab multi 12
30
ba75371e03
11/2009
4.3.4
AutoRead AR (Drift control)
The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of the
measured values.
Criteria
With identical measurement conditions, the following applies:
1
Call up the measuring mode with
▲> <▼> .
2
Activate the AutoRead function with
sured value is frozen (hold function).
3
Start AutoRead with
AR flashes until a stable measured value is reached. This mea-
sured value is transmitted to the interface.
4
If necessary, start the next AutoRead measurement with
5
To terminate the AutoRead function: Press the
8
LoBat
Time
Baud
Day.Month No.
Ident
Year
Sal
%
mbar
1
S
RCL
Store
Auto
Cal
°
F
1/cm
/
K
%
TDS
UpH
Sal
mV/pH
S/
m
cm
cm
M
Lin
Tref25 Tref20
nLF
AutoCal DIN
AutoCal TEC
6
7
1
0 8
O
°
C
mg/l
TP
ARng
AR
7
Measuring mode
Reproducibility
Response time
D. O. concentration
Better than 0.05 mg/l
> 10 seconds
D. O. saturation index
Better than 0.6 %
> 10 seconds