VEGA VVO VEGA Visual Operating User Manual

Page 9

background image

VEGA Visual Operating

9

Start VVO

In the VVO there are three different
user levels. As

Operator you can make

no modifications on the connected
instruments. Just observation and
monitoring are possible. At

Mainte-

nance configuration of a measurement
loop is not possible. Other functions
such as e.g. adjustment or editing of
linearisation curves are possible. In the
planning level all functions except the
service adjustments are accessible. In
this operating instruction we always
assume that you are in the planning
level. Click to

Planning and then click to

OK. As operator you are not asked for
a user identification. With maintenance

you are asked for a user identification. As planning engineer in addition a
password should be entered. Preadjusted for name and password is
"VEGA“. Later (with started VVO) the passwords can be modified. It is also
possible to deactivate the password inquiry. Click to

OK after you have

entered the name and the password.

This window shows that VVO gets in contact with the connected
instruments.

If this message appears, the failures can be the following:
-

Failure 1: Interface cable or VEGACONNECT 2 not connected

-

Failure 2: Interface cable or VEGACONNECT 2 connected to the
wrong standard interface

-

Failure 3: You have installed an older VVO-version (< 2.20). Use
the older interface converter VEGACONNECT.

Immediately remove, if possible, the failure and click to

Repeat. If the

failure cannot be removed, click to

Cancel and quit VVO. When the

failure is removed, start VVO again.

Choose

Direct cable connection (the operation via

modem is described in the operating instruction "Re-
mote parameter adjustment").