beautypg.com

Spare part list, Converter cg – M&C TechGroup CG-2H-W_-2MH-W Operator's manual User Manual

Page 28

background image

28

Gas sampling and conditioning technology

9-0.2-ME

LED‟s do not light up
Valves do not switch over
Temperature controller out
of order

No mains power




Sub-D-plug not inserted in
socket X2





F1/2 fuses defective

Mains „ON‟
OK.
Check that mains cable fits properly
(X1);
OK.
Check whether sub-D-plug is present
and is properly plugged in;
OK.
Check whether bridge in sub-D-plug
exists (s. Fig. 6);
OK.
Check fuses and replace if necessary;

Converter does not heat up
LED K1 off




Heater defective




Temperature controller
defective

Solid-state relay defective
Sensor short

Measure voltage at terminals X4/2 and
3;
OK.
Replace heater;
Not OK.
Measure voltage at terminal X4/6 and
7;
Voltage < 8V DC?
Check controller according to operat-
ing instructions;
Voltage > 8V DC?
Replace solid-state relay;
Change sensor;

Valves do not switch over
LED‟s do not light up

No mains supply (see above)
Sub-D-plug not inserted into
socket X2 (see above)
Internal circuit;
No solder link 1-6 in sub-D-
plug
External circuit;
Error in external control

see above

see above

Check sub-D-plug and if necessary
solder link;

Check external control;

Valves do not switch over
LED‟s light up

Valve defective

Check that valves function

No sample gas flow

Valves defective (see above)
Gas sample lines
blocked or leaking

See above
Check gas sample lines;
Check for gas-tightness

No conversion

Cartridge spent
Cartridge does not heat up
(see above)
No sample gas flow (see
above)

Change cartridge (17.1)
See above


See above



20

SPARE PART LIST

Wear, tear and replacement part requirements depend on specific operating conditions.
The recommended quantities are based on experience and are not binding.


Converter CG...