Gas sampling and conditioning technology

9-0.2-ME

Heating up

Heating flashes

Ramp function

Sensor control

 Set-point display, green
 PGM key

- to select the parameters

 Decrement key

- to alter values

 Increment key

- to alter values

 EXIT key

- to quit the levels

 LED for ramp/programme function, green
 LED for status indication, yellow

- outputs 1 to 3


18.2

OPERATING STATE OF THE TEMPERATURE CONTROLLER

The

operating state is displayed via the LED‟s 1 to 3 like:

Operating state

1

2

3

Heating up

off

on

blinking

Normal

heating on

on

on

on

heating off

on

on

off

Alarm

X

off

X

















In case of a high or low temperature alarm, the controller passes automatically into the state of self-
locking heater switch off.

18.3

CHANGING THE PARAMETERS OF THE TEMPERATURE REGULATOR

The programming of the regulator is made on different levels.
All important adjustments of the converter are listed in the user level and can be modified after remov-
ing the key locking.

In order to remove the locking, please proceed as follows:

Setting value reached


Operating temperature
or setting value reached